These Class I safety cabinets provide reliable containment for low-risk biological samples, safeguarding both the operator and the surrounding environment. Available in four sizes (1100, 1300, 1600, 1900), they feature robust construction and advanced control systems for laboratory and research use.
Dimensions & Capacity
-
Model Variants (External WxDxH):
-
BSC-I-1100: 1050 × 710 × 1850 mm
-
BSC-I-1300: 1320 × 710 × 1850 mm
-
BSC-I-1600: 1625 × 710 × 1850 mm
-
BSC-I-1900: 1930 × 710 × 1850 mm
-
-
Working Chamber (Internal WxDxH):
-
Ranges from 950 × 615 × 625 mm up to 1830 × 615 × 625 mm
-
-
Work Surface Height: 750 mm
Airflow & Filtration
-
Air Velocity: Adjustable between 0.2–0.5 m/s
-
HEPA Filtration: Single high-efficiency filter (> 99.995% at 0.3 μm)
-
Cleanliness Rating: ISO Class 5 (Class 100)
-
Back-wall Perforations: Promote even airflow and eliminate dead-air zones
Control & Safety Systems
-
Controller: Microprocessor with LCD showing fan status, filter life, UV life, and temperature
-
Alarms: Audio/visual alerts for filter replacement, UV lamp, sash position, and system errors
-
Interlocks: Prevent fan or UV operation when the front sash is open
-
Backup Memory: Retains settings after power interruptions
Lighting & UV Sterilization
-
Work Light: LED panel rated ≥ 900 Lux
-
UV Lamp: 253.7 nm germicidal output, timer-controlled (0–999 min)
-
Glass Panels: Front and sides use UV-resistant tempered glass for clear viewing and protection
Construction & Ergonomics
-
Housing: Electrogalvanized steel with antimicrobial powder coat
-
Work Surface: Angled, rounded-edge 304 stainless steel for easy cleaning
-
View Head: 45° inclined sliding sash with ergonomic handle
Power & Utilities
-
Supply: AC 220 V ±10% (50/60 Hz) or 110 V ±10% (60 Hz)
-
UV/Light Consumption: Between 155 W and 300 W depending on model
-
Integrated Sockets: Two IP54-rated outlets for small instruments
-
Optional Taps: Water or gas fittings available
Accessories & Packaging
-
Standard Inclusions: Base stand, LED work lamp, UV lamp, waterproof sockets
-
Optional Add-Ons: Integrated water or gas taps
-
Carton Size & Weight: Varies by model (e.g., 1190 × 990 × 1325 mm at 225 kg for the 1100 model)
Applications
-
Microbiological work with non-pathogenic samples
-
Forensic sample preparation
-
Aseptic compounding of non-hazardous products